UNI-T UT 4000 Series Manuel d'utilisateur Page 49

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 99
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 48
38
UNI-T
®
UT4000 User Manual
X-Y Mode
This mode is suitable for CH1 and CH2 only. After selecting the X-Y display mode, the
horizontal axis will display the CH1 signal, while the vertical axis will display the CH2
signal.
Figure 2-22 Waveform display in X-Y mode
Caution :
In the normal X-Y mode, you can adjust waveform amplitude with VOLTS/DIV of the
two channels. Adjusting the SEC/DIV time base knob will result in a lissajous figure of
better display quality.
The following functions are disabled in the X-Y display mode :
Automatic measurement mode
Cursor measurement mode
Reference or maths waveform
Vector display type
Horizontal [POSITION] knob
Trigger control
Key points :
Display Type : Vector display fills the space between adjacent sampling points in display.
The dot mode displays only the sampling points.
Refresh Rate : Refresh rate is the number of times the digital storage oscilloscope
refreshes the waveform display per second. The refreshing speed affects the capability to
observe signal movements speedily.
Vue de la page 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 98 99

Commentaires sur ces manuels

Pas de commentaire