4. Calibration
page 29
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
4. Calibration
A calibration is recommended to be performed regularly according to applied quality standards.
Go to the main menu and press F4: Calibration.
In the Calibration Mode menu you can select:
F1- to show the last performed calibration
F2- to show the history of all calibration data (can be deleted with F4)
F3- to perform a span check
F4- to perform a new calibration.
4.1 Show actual calibration
Press F1 in the Calibration Mode menu. The actual calibration parameters are displayed on the
screen.
By pressing the arrow-key one can have a look on the calibration data and on the calibration
curve.